  • Analytical Chemistry and Spectroscopy  (1)
  • Automatic control of induced paralysis  (1)
  • 1
    ISSN: 1573-2614
    Keywords: Automatic control of induced paralysis ; in vivo automatic control of paralysis ; vecuronium ; automatic infusion of anesthetic agents
    Source: Springer Online Journal Archives 1860-2000
    Topics: Computer Science , Medicine
    Notes: Abstract Objective. Reliable closed loop infusion systems for regulating paralysis level can be a great convenience to the anesthesiologists in automating their task. This paper describes the in vivo performance evaluation of a self-tuning controller that is designed to accommodate large varations in patient drug sensitivity, drug action delays and environmental interfering noise. Methods. The infusion system was evaluated in six adult mongrel dogs. Following the manual induction of paralysis by an anesthesiologist, the controller regulated the infusion of vecuronium to maintain a desired level of paralysis. The integrated EMG response of the hypothenar muscle to a train-of-four stimulation of the ulnar nerve quantified the depth of paralysis. The controller's robustness was tested by contaminating the sensed twitch signal with electrocautery noise and electrode disconnection. Results. The controller reached the initial level of paralysis of 100% in about 4.0 minutes and arrived at the desired level of 90% with an overshoot of 6.38% (±6.82). It maintained the desired level of paralysis with a 2.04% (±1.20) mean offset at 90% and 0.4% (±0.5) mean offset at 80% steady state level, respectively. The mean infusion rate to sustain 90% and 80% paralysis were 2.70 (±2.05) and 2.15 (±2.57) ((mg/kg)/min), respectively. Conclusions. The system adapted to a large variation in the sample subject drug sensitivity. It remained stable despite large amplitude disturbances and maintained the paralysis at the desired level following the removal of the disturbances.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
  • 2
    Electronic Resource
    Electronic Resource
    Chichester : Wiley-Blackwell
    Organic Magnetic Resonance 24 (1986), S. 308-311 
    ISSN: 0749-1581
    Keywords: Chemistry ; Analytical Chemistry and Spectroscopy
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Chemistry and Pharmacology
    Notes: Carbon-13 NMR data of [5]metacyclophane (1a), its 8,11-dihalosubstituted derivatives 1b-1d and of some model compounds are reported and discussed. A remarkable downfield shift is observed for the aromatic carbon between the bridging carbons; this is tentatively ascribed to long-range effects of the bridge and to the bending of the aromatic ring.
    Additional Material: 1 Ill.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability