---A new ECD, the HP 6890 Series Micro-ECD, was designed to address inherent deficiencies in classical electron capture detectors (ECD), especially with respect to sensitivity, linearity, dynamic range, and ruggedness. Several novel technologies were incorporated in the totally new design and were refined through practical testing and user feedback. Validation of the micro-ECD performance was accomplished through side-by-side testing of the Micro-ECD with previous ECDs following US EPA Contract Laboratory Program methods for pesticides and PCBs. In addition, extensive interviews were conducted with early users of the Micro-ECD who also had experience with other designs running a variety of ECD methods. The design and resulting performance improvements are described.