ISSN:
1741-0444
Keywords:
Analysis
;
Autocorrelation
;
Defibrillation
;
Performance
;
Regression
;
Signal
;
Ventricular fibrillation
Source:
Springer Online Journal Archives 1860-2000
Topics:
Biology
,
Chemistry and Pharmacology
,
Medicine
Notes:
Abstract The paper investigates quantitative differences in the signal characteristics of ventricular fibrillation (VF) and other cardiac arrhythmias. The analysis procedure comprises two steps: calculation of a short-term autocorrelation function (ACF) followed by a regression test on a plot of peak magnitudes of the ACF against lag values (the ACF/lag plot). We detect VF by testing the hypothesis that the ACF/lag plot of VF does not pass a linear regression test. Analysis of 31 separate episodes (of VF and other ventricular arrhythmias), each comprising three successive segments of 1·5s each produced the following results: (1) 100 per cent sensitivity (Se), 62 per cent specificity (Sp) and 74 per cent test efficiency (TE) after analysis of the first segment; (2) 100 per cent Se, 86 per cent Sp and 90 per cent TE after the second segment; and (3) 100 per cent Se, 100 per cent Sp and 100 per cent TE after the third segment. This method quantifies the notion that VF signals are nonperiodic with a random amplitude distribution, whereas ventricular tachycardia (VT) signals are usually periodic with more uniform amplitude distributions. Accurate discrimination and identification of VF can be very important in intensive-care settings, as well as in the design of automatic cardioverters and defibrillators.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF02447420
Permalink