Electronic Resource
Oxford, UK
:
Blackwell Publishing Ltd
Anaesthesia
47 (1992), S. 0
ISSN:
1365-2044
Source:
Blackwell Publishing Journal Backfiles 1879-2005
Topics:
Medicine
Notes:
Measurements from the upgraded Finapres 2300e continuous noninvasive blood pressure monitor, the Finapres 2300 and Colin oscillometric noninvasive blood pressure monitor were compared with invasive arterial line blood pressure readings. Fifteen young Chinese patients undergoing elective spinal surgery of more than 2 h duration had contemporaneous blood pressure measurements digitally recorded every minute. Data were analysed using methods described by Bland and Altman to assess agreement of noninvasive devices with an arterial line. Results showed that although the Finapres 2300e was significantly more accurate than the Finapres 2300 and Colin noninvasive blood pressure monitors it could not be recommended as a substitute for continuous arterial line blood pressure monitoring. Both Finapres devices demonstrated reductions in accuracy related to time (drift) and over-read diastolic and mean pressures by 5–8 mmHg throughout the range of mean arterial line pressures (bias). The Colin was consistently less accurate than the Finapress monitors and performed worst at low mean arterial line pressures.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1111/j.1365-2044.1992.tb02396.x
Permalink
Library |
Location |
Call Number |
Volume/Issue/Year |
Availability |