Summary Intra-arterial measurement is considered the gold standard for continuous, beat-to-beat arterial blood pressure monitoring. However, arterial cannulation can be difficult and may cause complications such as thrombosis and ischaemia. Recently, a tonometric system, the Colin CBM-7000 has been developed for noninvasive beat-to-beat measurement of arterial blood pressure from the radial artery. We assessed the level of agreement between the CBM-7000 and invasive radial artery measurements in 15 patients on a neuro-intensive care unit. Agreement of systolic, diastolic and mean arterial pressure values was limited, with ≈ 34% of mean arterial pressures differing by over 10 mmHg. In many cases, this was due to a downward drift of the noninvasive measurements over time. Furthermore, there was a tendency to underestimate low pressures and overestimate high pressures. In our opinion, the Colin CBM-7000 cannot be recommended for continuous blood pressure monitoring in the intensive care setting.