Fluid delivery from four types of commercially available 50-ml syringes was measured using an electronic balance at an infusion rate of 1 ml.h−1. Retrograde aspiration volume and zero-drug delivery time were recorded after lowering the syringe pump by 50 cm. Syringe compliance was calculated from the volume of bolus released after occlusion at 100 mmHg. Zero-drug delivery times differed significantly between syringes, ranging from [mean (SD)] 3.26 (0.40) min to 6.38 (0.56) min (F = 55.5, d.f. = 3/20, p 〈 0.0001). Syringe compliance correlated well with aspiration volume (Pearson r2 = 0.92, p 〈 0.001) and zero-drug delivery time (r2 = 0.90, p 〈 0.001). Syringe design affected the internal syringe compliance. All syringes were associated with potentially relevant zero-drug delivery times after moderate vertical displacement. To minimise this risk, vertical displacement of syringe pumps delivering highly vasoactive drugs should be avoided.