Electronic Resource
Chester
:
International Union of Crystallography (IUCr)
Journal of synchrotron radiation
7 (2000), S. 18-21
ISSN:
1600-5775
Source:
Crystallography Journals Online : IUCR Backfile Archive 1948-2001
Topics:
Geosciences
,
Physics
Notes:
A complementary pair of rapid methods for the energy calibration of solid-state detectors have been developed. Each method requires only a single measurement from either (i) a glass sample containing elements chosen to produce fluorescence lines over a suitable energy range when exposed to a polychromatic beam of X-rays, or (ii) a powder diffraction standard in the presence of diffraction slits. The fluorescent glass method has the advantage of allowing simultaneous energy calibration of a number of detectors without requiring diffraction slits. There is the potential for the glass material to be incorporated into virtually any sample holder to allow a continuous in situ calibration. Complementary observations of a powder diffraction standard allow simultaneous calibration of diffraction 2θ and monitoring of the detector for drift.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1107/S0909049599014636
Permalink
Library |
Location |
Call Number |
Volume/Issue/Year |
Availability |