Average velocity
;
Fast GC
;
High pressure drop
;
Plate height
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
Typically, fast GC requires high column pressure drop. Under that condition, known relations do not provide a complete description of dependence of [apparent] plate height, ĥ, on average velocity, u, of carrier gas.A complete descriptions for ĥ vs. u is derived and analyzed. It is shown that, under the high pressure drop, ĥ = B/u2 + C1 u2 + C2 u. This is substantially different from the Van Deemter equation ĥ = B/u + Cu for the low pressure drop. Analytical expressions for the minimum plate height and for the respective optimum u under the high pressure drop are derived and analyzed.