    Archives of orthopaedic and trauma surgery 112 (1992), S. 15-17 
    ISSN: 1434-3916
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Notes: Summary The excimer laser was thought to be an appropriate tool for the removal of bone cement without damaging the bone. However, due to its low ablation rate, its clinical use in total hip revision arthroplasty proved to be impossible. This experimental study was designed to evaluate the maximal ablation rate by adjusting the laser's parameters. Energy density, frequency, pulse duration, radiation area, quantity of pulses, and environmental conditions were varied in the experimental setup. Even with the best set of parameters the excimer laser was about ten times slower than, e.g., the carbon dioxide laser. The removal of 10 g bone cement takes about 1 h. Thus, complete cement removal by means of the excimer laser alone is not possible. However, selective application of the excimer laser in combination with other techniques could be discussed.