ISSN:
1612-1112
Keywords:
Column liquid chromatography
;
Membrane-extraction disks
;
On-line trace enrichment
;
Polar pollutants
;
Surface water
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Notes:
Summary An on-line trace-enrichment system, using a bifunctional membrane extraction-disk cartridge, has been combined with a column liquid chromatography separation for the simultaneous determination of basic, neutral and acidic pollutants in surface water. The enrichment device consisted of a specially constructed holder containing both C-18 and cation-exchange disks. The holder can contain up to 25 disks of 0.5 mm thickness and 4.6 mm diameter. Before trace enrichment of 20 ml of surface water (pH 3.0), calcium ions were removed from the sample by means of an oxalic acid precipitation. Desorption of the cartridge was at elevated temperature using reversed-phase gradient elution; detection was performed with a diode-array UV absorbance detector. The detection limits for the test compounds in surface water are typically 0.5–2 μg/l; the calibration graphs are linear from the lower limit of determination up to 50 μg/l.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF02274506
Permalink