Description / Table of Contents:
Summary A study of the accuracy of the tpH electrode was performed in 10 cats. Hypoxia and acidosis were produced by ventilatory means and bicarbonate infusions were used to elevate the pH. Tissue pH correlated well with arterial (r = 0.9) within the physiological rates of change of pH. Prolonged tissue acidosis lead to eventual death. The correct and stable mechanical fixation of the pH electrode is a critical factor in assuring accurate results. The ROCHE tpH electrode accurately reflects arterial pH, and a correlation coefficient of at least 0.85 should be attainable in the human fetus.
Notes:
Zusammenfassung Bei zehn Katzen wurde die Meßgenauigkeit der tpH-Elektrode überprüft. Hypoxie und Acidosis wurde durch Änderung der Beatmung erzeugt, während durch Bikarbonat-Infusion die pH-Werte angehoben wurden. Innerhalb physiologischer Grenzen der pH-Wert-Änderungen zeigte sich eine gute Korrelation (r = 0,9) zwischen dem Gewebe-pH-Wert und den arteriellen ph-Werten. Eine Verlängerung der Acidose im Gewebe führte schließlich zum Tode. Die korrekte und stabile Anbringung der pH-Elektrode ist ein kritischer Faktor bei den Bemühungen um genaue Meßergebnisse. Die Gewebe-pH-Elektrode von ROCHE gibt exakt die arteriellen pH-Werte wieder und ein Korrelationskoeffizient von mindestens 0,85 dürfte beim menschlichen Feten erreichbar sein.