Abstract The turbulence generated downstream from an aortic Ionescu-Shiley bioprosthesis has been investigated in vitro with both steady and pulsatile flow; Instantaneous point velocities were measured using laser-Doppler anemometry (LDA) at numerous preselected locations in the flow. The mean and RMS velocities from these data at each location were then used to estimate the laminar and turbulent shear in the bulk flow as a function of radial position on a cross-section of the flow system downstream from the mounted prosthesis. Estimated total shear stresses were found in the bulk flow that were of sufficient magnitude to possibly cause haemolysis and initiate platelet chemical-release reactions. For steady flow and at peak pulsatile flow, maximum total shear stresses were estimated to be 120 N m−2 and 100 N m−2, respectively, over more than 5 per cent of the flow cross-section. The spatial distribution of the elevated shear stresses correlates well with the valve superstructure. It is concluded that these elevated stresses are a direct consequence of the notable flow constriction generated by the valve’s fully opened leaflets.