ISSN:
1618-2650
Keywords:
Best. von Cadmium
;
Spektralphotometrie, Atomabsorption
;
flammenlos, Störungen
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Description / Table of Contents:
Summary Interferences in the flameless determination of cadmium using the HGA 72 were studied for HCl, HNO3, H2SO4, HClO4 and HF and found to be depending on different thermal pretreatment. The interferences of cations were much smaller than for anions. The results are discussed with respect to thermal stability as well as atomization rate of the compositions under investigation. Special emphasis is put upon the presence of interactions between interferents, which in some cases can be used to eliminate interferences.
Notes:
Zusammenfassung Interferenzen, die bei der flammenlosen Bestimmung von Cadmium in der HGA 72 auftreten, wurden bei HCl, HNO3, H2SO4, H3PO4, HClO4 und HF studiert; die AbhÄngigkeit ihres Auftretens von der thermischen Vorbehandlung wird gezeigt. Interferenzen von Kationen sind kleiner. Die Ergebnisse werden im Hinblick auf die thermische StabilitÄt und die Atomisierungsgeschwindigkeit der Proben diskutiert. Besonders betont wird die Gegenwart von Wechselwirkungen zwischen Störungen, die manchmal auch benützt werden können, um Interferenzen auszuschalten.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF00539072