    Electronic Resource
    Electronic Resource
    Springer
    Journal of thermal analysis and calorimetry 49 (1997), S. 343-349 
    ISSN: 1572-8943
    Keywords: double melting peak ; DSC ; fiber ; polyethylene
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Abstract The multiple melting peaks observed on differential scanning calorimetry (DSC) of ultrahigh molar-mass polyethylene fibers (UHMMPE) are analyzed as a function of sample mass. Using modern DSC capable of recognizing single fibers of microgram size, it is shown that the multiple peaks are in part or completely due to sample packing. Loosely packed fibers fill the entire volume of the pan with rather large thermal resistance to heat flow. On melting, the fibers contract and flow to collect ultimately at the bottom of the pan. This process seems to be able to cause an artifact of multistage melting dependent on the properties of the fibers. A method is proposed to greatly reduce, or even eliminate, errors of this type. The crucial elements of the analysis of melting behavior and melting temperature are decreasing the sample size and packing the individual fibers in a proper geometry, or to introduce inert media to enhance heat transport.
    Type of Medium: Electronic Resource