    Keywords: Radioactive Wash-out Technique ; Renal Blood Flow ; Electromagnetic Flowmeter ; Isolated Kidney ; Inert Gas Desaturation ; Isotopen-Auswaschtechnik ; Nierendurchblutung ; Elektromagnetischer Strömungsmesser ; Isolierte Niere
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Notes: Summary By use of an electromagnetic flowmeter and the inert gas desaturation method 43 comparative measurements of the blood flow were performed in 6 isolated dog kidneys. For the measurements by use of the latter method krypton-85 and xenon-133 were employed, for a fast, and respectively, a slow injection of the indicator into the renal artery. Calculation of the blood flow from the wash-out curves was made by use of the formulas advanced by Kety (1951) and by Zierler (1965). A good accordance was found between the flowmeter values and those obtained with the desaturation methods. Recirculation and resorption of indicator in fat tissue surrounding the kidney explain that previous examinations in kidneys in situ have resulted in very low perfusion values, when Zierler's formula has been applied. Extrapolation of the final slope to obtain the area of the curve may be a major source of error by calculation of the blood flow with Zierler's formula.