Library

feed icon rss

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    ISSN: 1432-0932
    Keywords: Pedicle screw ; Placement ; Accuracy ; Plain radiographs ; CT reconstruction
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Notes: Abstract The position of 119 pedicle screws was assessed on plain antero-posterior and lateral radiographs taken immediately post-operatively and at 3 months' follow-up. The readings of five independent observers were compared with the “gold standard” of CT reconstructions. The position of only 41 % of implants (range 14%–56%) was assessed correctly on the plain radiographs (47% on follow-up films). Two-thirds of CT-detectable perforations were missed. As shown with perforations of the anterior cortex, delectability increased significantly with magnitude of perforation. No specifically difficult anatomic level or direction of malplacement could be identified. Interobserver variation was considerable. Plain radiographs were shown to be of limited use in assessing the position of pedicle screws.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability