Library

feed icon rss

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
Filter
  • Prediction of retention times  (1)
  • 1
    Electronic Resource
    Electronic Resource
    Springer
    Chromatographia 38 (1994), S. 715-722 
    ISSN: 1612-1112
    Keywords: Reversed phase ion-pair chromatography ; LC-optimization ; Chemometrics ; Prediction of retention times
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Summary A strategy for general optimization of the mobile phase in LC-systems is described in this study. The influence on column efficiencies and capacity factors of three uncharged modifiers, two charged modifiers, pH and column temperature was evaluated with experimental design and multivariate analyses. A factorial design was used and the experimental results demonstrated that the uncharged modifiers and the pH were important for the capacity factors while column temperature and pH were important for the column efficiency. Two different stationary phase materials were used and compared, one was a traditional silica based material and the other a silica covered with a polymer. Both were reversed phase materials. A partial least squares model obtained from the results from the polymer column, Chromspher B, was used to predict retention data.