Abstract The European Gas Research Group (GERG; Groupe Européen de Recherches Gazières) initiated a round-robin test of six Z-meters manufactured by Desgranges et Huot, a Burnett apparatus, and an interferometric device to back up the pVT data of the Z-meters. Two gas mixtures were measured. One mixture contained 49.7 mol% of methane and 50.3 mol% of nitrogen; the second mixture 81.3 mol% of methane, 16.4 mol% of ethane, and 2.3 mol% of propane. The test temperatures were mainly 280 and 300 K for the first mixture and 290 and 320 K for the second mixture. The maximum pressures were 8 MPa for Z-meters and 12 MPa for the Burnett apparatus and the grating interferometer. The experimental compressibility factors Z of the six Z-meters are generally in agreement within ±0.05%. The agreement with the reference data from the Burnett apparatus and the refractive index measurements is also within ± 0.05%. Only two isotherms of the binary mixtures differ by about 0.1% from the other data. Recent natural gas measurements show substantially the same results.