    Keywords: Gas chromatography ; Gas analysis ; Determination of gas fractions ; Accuracy of the Scholander method
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Notes: Summary The volume fraction of gas mixtures may be measured gas chromatographically by application of the 100% method, provided they can be separated. In case of gases it is not necessary to calibrate with another method. The specific character of the detector signals can be eliminated by measurements with pure components. The procedure is compared with the Scholander method. In case of O2/N2 mixtures the results agree within reproducibility, but in case of O2/CO2/N2 mixtures a systematic deviation of the CO2 fractions to smaller values is found as concerns the Scholander method. A gas mixture with 5.6 Vol% CO2 shows a difference of approximately 0.3 Vol%. The gas chromatographic method is almost 100 times as sensitive as the Scholander method. High fractions may be measured without problems and the method is not restricted to O2 and CO2.