ISSN:
1432-1084
Keywords:
Key words: Stress urinary incontinence – Intra-vaginal device – MRI – Dynamic MRI
Source:
Springer Online Journal Archives 1860-2000
Topics:
Medicine
Notes:
Abstract. The aim of this study was to describe by MRI, in dynamic conditions at rest and straining, the anatomical modifications induced by a commercially available intravaginal device (IVD) aimed at relieving female stress urinary incontinence. Ten female patients complaining of stress urinary incontinence (SUI) had pelvic MRI with static and dynamic sequences, without and with a self-inserted IVD. We studied positions and angulations of the IVD in the pelvis. Paired t-test allowed comparisons of: position of the bladder neck; urethral angulation with the pubis axis; position of the urethra; and posterior urethro-vesical angle (PUVA) without and with IVD. At rest, in ten of ten cases IVD laid cranial to the pubo-rectal muscle; with an average angulation of 95 ± 10 ° with the pubis axis, laterally tilted in three of ten cases. In maximum straining with IVD bladder neck descent was lower by an average of 5.2 ± 3.1 mm (p = 0.001), pubo-urethral angle opening was smaller by an average of 22 ± 20 °(p = 0.015), and bladder neck to pubis distance was shorter by an average of 5.7 ± 4 mm. Posterior urethro-vesical angle was not significantly modified. Dynamic MRI allowed a non-invasive assessment of the mode of action of an IVD. The main modifications were a support of the bladder base and bladder neck, with a superior displacement of the urethra toward the pubis.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/s003300051028
Permalink