Electronic Resource
350 Main Street , Malden , MA 02148-5018 , USA and 9600 Garsington Road , Oxford OX4 2DQ , UK .
:
Blackwell Science Inc
Pacing and clinical electrophysiology
28 (2005), S. 0
ISSN:
1540-8159
Source:
Blackwell Publishing Journal Backfiles 1879-2005
Topics:
Medicine
Notes:
This is a report on a patient with an implanted cardioverter defibrillator (ICD) who intentionally underwent magnetic resonance imaging (MRI) of a malignant brain tumor. To avoid inadequate detection of ventricular tachycardia (VT) or ventricular fibrillation (VF), the ICD was inactivated by programming the VT-detection and VT/VF-therapy status off. The patient came through the protocol safely and without any difficulty or discomfort. There was no arrhythmic event. MRI affected neither programmed data nor the function of the ICD system.