ISSN:
0741-0581
Schlagwort(e):
Ultramicrotomy
;
Serial sectioning
;
Electronmicroscopy
;
Seria reconstruction
;
Life and Medical Sciences
;
Cell & Developmental Biology
Quelle:
Wiley InterScience Backfile Collection 1832-2000
Thema:
Allgemeine Naturwissenschaft
Notizen:
The process of serial sectioning for electron microscopy has been refined such that loss of thin sections is kept below 0.1% and the series is continued at will. The method relies on microscopic control of all manipulative steps, Formvar casting on plate glass for coated slot grids, coating of the block with contact cement for reliable ribboning, pickup by a one-step method with grid support in the diamond knife trough, staining in LKB grid holders, gentle treatment of grids in the electron microscope, and a slight modification to the microscope for safe grid withdrawal. The results are particularly applicable to the reconstruction of neuronal microcircuits and larger volumes of neuropil.
Zusätzliches Material:
12 Ill.
Materialart:
Digitale Medien
URL:
http://dx.doi.org/10.1002/jemt.1060010407