    Keywords: Miniaturized HPLC ; Post-Column Extraction Detection ; Segmented-Flow Open-Tubular Reactors ; Membrane Phase Separator ; Anti-Cancer Drugs
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Abstract An on-line, post-column, extraction system suitable for use with miniaturized HPLC (0.7–1 mm i.d. columns) is described. The system consists of a 30° angle segmenting tee, a PTFE extraction coil of variable length and a membrane phase separator. The internal volume of the phase separator is less than 1 μl. The total variance contribution of the system is between 1 and 6 sec2. The determination of the anti-cancer drugs VP 16 and VM 26 in urine by means of reversed-phase HPLC and fluorescence detection, using both on-column preconcentration and the present miniaturized post-column extraction system is presented as an application.