ISSN:
1618-2650
Schlagwort(e):
Best. von Lactat, Lactatdehydrogenase
;
Biamperometrie
;
Hexacyanoferrat(II), Durchfluß, Kohlenstoffelektrode
Quelle:
Springer Online Journal Archives 1860-2000
Thema:
Chemie und Pharmazie
Beschreibung / Inhaltsverzeichnis:
Summary Open tubular carbon electrodes were used to monitor the biamperometric current of hexacyanoferrate(II) ion produced by the LDH catalyzed reaction between hexacyanoferrate(III) ion and lactate, and that produced from the diaphorase catalyzed reaction between hexacyanoferrate(III) ion and NADH, product from the LDH catalyzed reaction between lactate and NAD. Reagents and samples were mixed in a flow stream. Reaction takes place between the mixing point and the electrodes, and measurements are taken after a constant time interval. Lactate and LDH standards in serum control solutions were measured using the NAD dependent reaction.
Notizen:
Zusammenfassung Kohlenstoff-Röhrenelektroden wurden zur Messung des biamperometrischen Stromes des Hexacyanoferrat(II)-ions eingesetzt, das aus der LDH-katalysierten Reaktion zwischen Hexacyanoferrat(III) und Lactat bzw. der Diaphorase-katalysierten Reaktion zwischen Hexacyanoferrat(III) und NADH (Produkt der LDH-katalysierten Reaktion zwischen Lactat und NAD) stammte. Die Reagentien wurden mit den Proben im Durchfluß vermischt. Die Reaktion fand zwischen der Mischungsstelle und den Elektroden statt und die Messungen wurden nach einem Bestimmten Zeitintervall durchgeführt. Mit Hilfe der NAD-abhängigen Reaktion wurden Lactat- und LDH-Standards in Serumkontrollösungen bestimmt.
Materialart:
Digitale Medien
URL:
http://dx.doi.org/10.1007/BF01171503
Permalink