ISSN:
0173-0835
Schlagwort(e):
Polyacrylamide gel
;
Matrix-assisted laser desorption
;
ionization
;
Membranes
;
Mass spectrometry
;
Blotting
;
Chemistry
;
Biochemistry and Biotechnology
Quelle:
Wiley InterScience Backfile Collection 1832-2000
Thema:
Biologie
,
Chemie und Pharmazie
Notizen:
Matrix-assisted laser desorption ionization (MALDI) mass spectra have been obtained directly from thin-layer isoelectric focusing (IEF) gels with as little as 700 femtomoles of α- and β-chain bovine hemoglobin and bovine carbonic anhydrase, and 2 picomoles of bovine trypsinogen, soybean trypsin inhibitor, and bovine serum albumin all loaded onto a single lane. By soaking the gel in a matrix solution, matrix was deposited over the entire gel surface, allowing MALDI scanning down complete lanes of the one-dimensional gel. As long as matrix crystals were deposited finely on the surface of the gel, time-lag focusing techniques were capable of ameliorating some of the mass accuracy limitations inherent in desorbing from uneven insulator surfaces with external calibration. Eleven measurements on the 5 kDa α-subunit proteins of lentil lectin measured over the course of 1 h and referenced to a single calibration yielded a standard deviation of 0.025%. Colloidal gold staining was found to be compatible with desorption directly from IEF and sodium dodecyl sulfate (SDS)-polyacrylamide gels. This direct approach simplifies the interface between gel electrophoresis and mass spectrometry dramatically, making the process more amenable to automation.
Zusätzliches Material:
6 Ill.
Materialart:
Digitale Medien
URL:
http://dx.doi.org/10.1002/elps.1150180312