    Journal of High Resolution Chromatography 5 (1982), S. 424-427 
    ISSN: 0935-6304
    Keywords: Liquid chromatography, HPLC ; Column switching ; Step gradient ; Immunosuppressive agent ; Determination in blood and plasma ; Deproteinization ; On-line sample clean-up ; Chemistry ; Analytical Chemistry and Spectroscopy
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Chemistry and Pharmacology
    Notes: A method for the determination of cyclosporin A in human whole blood and plasma is described which uses liquid chromatography with step gradient elution and a column switching technique. The chromatographic conditions chosen allow simple and rapid sample preparation, so that a result can be obtained within one hour. Blood and plasma are deproteinized with diluted methanol and an aliquot of the clear supernatant is directly injected. The detection limit for cyclosporin A is about 20 ng/ml starting from a 0.5 ml sample. The method is sensitive enough for monitoring the drug in the therapeutic range.
    Additional Material: 4 Ill.