ISSN:
1612-1112
Keywords:
Column liquid chromatography
;
Ibuproxam and Ibuprofen in plasma
;
Column switching
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Notes:
Summary A new high-performance liquid chromatographic method with column switching has been developed for the simultaneous determination of Ibuproxam and its metabolite Ibuprofen in plasma. The plasma samples were injected onto a LiChrosorb RP-8 precolumn after dilution with an internal standard solution in phosphoric acid. Polar plasma components were eluted with 0.05 M phosphoric acid. After valve switching, the concentrated drugs were eluted in the back-flush mode and separated on an Inertsil ODS-2-column with acetonitrile-methanolphosphate buffer as mobile phase. The method showed excellent precision, accuracy and speed with detection limit of 0.1 μg ml−1. Total analysis time per sample was less than 30 min and the coefficients of variation for intra and inter-assay were less than 3.5%. This method has been successfully applied to plasma samples from human volunteers after oral administration of Ibuproxam.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF02274112