Library

feed icon rss

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    Springer
    Medical & biological engineering & computing 17 (1979), S. 211-215 
    ISSN: 1741-0444
    Keywords: Hot-wire anemometer ; Respiratory-gas-flow measurements
    Source: Springer Online Journal Archives 1860-2000
    Topics: Biology , Chemistry and Pharmacology , Medicine
    Notes: Abstract The performance of a constant-temperature hot-wire anemometer for respiratory-gas-flow measurement has bee investigated. It is shown that the anemometer satisfies common demands with respect to accuracy, response time and to insensitivity to humidity and temperature variations. A calibration method is suggested that permits correction for variation of the flowmeter signal caused by changes of the gas composition. Finally, a procedure for linearisation of the flowmeter signal as a function of flow rate is presented.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗