    ISSN: 1432-1440
    Keywords: Insulin infusion pump ; Implantation ; CSII ; Intravenous, intraperitoneal insulin infusion
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Notes: Summary Five type I diabetic patients were followed prospectively during treatment with continuous subcutaneous insulin infusion by externally worn pumps and during the first 12 months after implantation of a remote-controlled insulin infusion device (ID1, Siemens AG). Stabilized insulin (Hoe 21 GH, Hoechst AG) was infused intravenously in two and intraperitoneally in three patients. Total observation time was 47.2 patient-months after implantation. Two devices had to be explanted prematurely, one because of a technical failure after 101 days, one due to a skin necrosis over the implant after 236 days. HbA1, frequency of hypoglycemia, total insulin dose, and basal rate infusion did not change after implantation. There was a reduction in the insulin antibodies 6 months after start of intravenous or intraperitoneal insulin delivery. Fasting plasma free insulin levels could be normalized only by intraperitoneal insulin infusion. Although a technical and a surgical problem was observed, our data show the successful implantation and clinical use of programmable dosing devices and stabilized insulin.