Summary This report documents a case of ball variance in a Smeloff-Cutter aortic prosthesis occurring 24 years after implantation. After episodes of embolic complications the patient died in acute shock. The silicone rubber ball showed several alterations including discoloration, grooving, cracking, swelling and subtotal fracture of the poppet. Terminal valvular malfunction was caused by complete thrombosis of the prosthesis. In most patients ball variance occurred during the first years after valve replacement; thus, the observed case is a very rare late complication of a ball-valve prosthesis.
Notizen:
Zusammenfassung Berichtet wird über einen Fall von Ballvarianz einer Smeloff-Cutter-Prothese in Aortenposition 24 Jahre nach Implantation. Nach vorausgegangenen embolischen Komplikationen verstarb der Patient im kardiogenen Schock. Der Silikon-Kautschuk-Ball der Prothese zeigte eine ausgeprägte Degeneration mit gelblicher Verfärbung, Rillen- und Rißbildung bei Schwellung und subtotaler Fraktur. Daraus resultierte eine terminale Prothesenfunktionsstörung mit kompletter Thrombose der Klappe. Ballvarianz als Komplikation trat bei der großen Mehrzahl der Patienten während der ersten Jahre nach Implantation auf; somit ist die beobachtete Komplikation ein sehr seltenes Ereignis nach Implantation einer Kugel-Käfig-Prothese.