    Springer
    Medical & biological engineering & computing 25 (1987), S. 45-50 
    ISSN: 1741-0444
    Keywords: Blood gas ; ISFET ; Lactate Ringer ; Patient monitoring ; pCO2 ; pH ; pO2
    Source: Springer Online Journal Archives 1860-2000
    Topics: Biology , Chemistry and Pharmacology , Medicine
    Notes: Abstract A system for the online, simultaneous and intermittent analysis of blood pH, pCO2 pO2 using two-way cyclic pumping of lactate Ringer as a baseline solution is described. The multi-gas sensor is composed of a pH-ISFET with a temperature sensor, a pH-ISFET-based pCO2 sensor and a Clark-type pO2 sensor, bundled in an epoxy rod, and held in place in a flow cell which is directly connectable to an indwelling cannula. The system is controlled by a microcomputer, which supervises pump operation, temperature-compensation of sensor outputs, one-point calibration using sampled blood analysis, and provides graphical display of monitored results. Results of arterial blood-gas monitoring for mongrel dogs are presented which demonstrate the reliability of the present system over extended periods.
    Type of Medium: Electronic Resource