Library

feed icon rss

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    Berlin : Wiley-Blackwell
    Acta Biotechnologica 7 (1987), S. 61-69 
    ISSN: 0138-4988
    Keywords: Life Sciences ; Life Sciences (general)
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Process Engineering, Biotechnology, Nutrition Technology
    Notes: A continuous flow system with an enzyme electrode and with a new type of measuring instrument, the BIOXY-Meter, is described as an example for the determination of glucose. The BIOXY-Meter combined with various biosensors may be used in biotechnology and in other fields for the solution of a number of analytical problems. Measuring principle is the reduction of oxygen. The peak of first derivation of the current-time-curve is proportional to oxygen consumption and to decrease of substrate concentration. Enzyme fixation with common adhesive and a device for automated saturation of measuring samples with air in the sampler are also described. With the apparatus 60 samples per hour are analyzed for glucose. The linear measuring range is 0.1 mMol to 1.5 mMol with a coefficient of variation of Vk = 0.93 per cent. The described type of enzyme electrode is stable for 2 to 12 weeks or 1,000 to 5,000 measurements.
    Additional Material: 4 Ill.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...