Library

feed icon rss

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    ISSN: 1572-817X
    Keywords: accuracy ; complex form factor f ; X-ray optics
    Source: Springer Online Journal Archives 1860-2000
    Topics: Electrical Engineering, Measurement and Control Technology , Physics
    Notes: Abstract Reliable knowledge of the complex X-ray form factor (Re(f) and f″) is required for many fields including crystallography, medical diagnosis and XAFS studies. However, there are discrepancies between theory and theory, experiment and experiment and theory and experiment of 10% and more, over central X-ray energies. Discrepancies exist for most elements, despite claimed experimental accuracies of 1%. This paper summarises the current variation between experimental and theoretical results, and outlines key issues for obtaining experimental accuracies of 1% in critical wavelength ranges for selected elements to address these issues. This paper critically surveys available experimental data for attenuation coefficients and suggests a procedure for obtaining significantly higher accuracy measurements in the future.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...