Keywords:
Gas chromatography
;
Open-tubular (capillary) columns
;
Comparison of liquid phases
;
Influence of temperature
;
Retention index data
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Notes:
Summary The influence of temperature and liquid phase film thickness of open-tubular (capillary) columns on the retention index values of hydrocarbons on methylsilicone liquid phases is discussed. Data obtained on methylsilicones and squalane are compared. Retention index values of 43 hydrocarbons between 40 and 70 °C on OV-101 liquid phase are listed.