Summary
After introductory remarks about the flow phenomena in the human body (separation, dead waters, stagnation regions) and the model simularity conditions, the flow behaviour in arterial bifurcations at pulsatile flow are considered; they are accompanied by the formation of dead waters and secondary flows and therefore they give the danger of atherosclerosis. Then the flow pattern at arterio-venous anastomosis is interpreted with regard to the danger of thrombogenesis.
At the selected example of ball prosthesis the flow pattern and the pressure loss of artificial heart valves (aortic as well as mitral valves) are described; the special behaviour of ball valves used as mitral valves is discussed critically.
Finally in a view of contribution to the question of the flow-induced blood damage, in extended series of basic tests on models (orifices and perforated discs) the rate of hemolysis was measured depending on the flow; the tests were completed by measurements at rolling pumps, atria of a artificial ventricle and other parts.
In order to try an explanation for the flow-induced hemolysis one starts from the influence of the hemodynamic shear stresses on the damage of the erythrocytes; hereby the effect of the frequently repeated passages of the red blood cells through a stenosis (e.g. regurgitation) should be taken into account.
Zusammenfassung
Nach einer kurzen Einführung in Strömungserscheinungen (Ablösung, Totwasser, Stagnationsbereiche) und in die Modellähnlichkeit werden die Vorgänge in Aderverzweigungen bei pulsierender Strömung mit der Bildung der durch Sekundärströmung bedingten Towasserbereiche als eine Voraussetzung der Atherosklerose betrachtet; anschließend werden Strömungsbilder bei arteriovenösen Anastomosen hinsichtlich ihrer Thrombogenesegefahr gedeutet. Strömungsverlauf und Druckverlust künstlicher Herzklappen werden am Beispiel von Ballprothesen besprochen; hierbei wird das beondere Verhalten der Ballprothesen als Mitralisklappe kritisch erörtert.
In längeren Reihen von Grundsatzversuchen mit Modellen (Lochscheiben, perforierte Scheiben) wurde durch Messung der Hämolyserate in Abhängigkeit vom Strömungszustand die Möglichkeit starker strömungsbedingter Blutschädigung nachgewiesen; die Versuche wurden durch Messungen an Rollenpumpen, Vorhöfen künstlicher Ventrikel u.a. ergänzt. Zur strömungstechnischen Deutung wird von der Wirkung der hämodynamischen Schubspannung auf die Hämolyse ausgegangen; als wesentlich wird auf den Effekt vielfach wiederholten Durchgangs der Erythrozyten durch hämolysebegünstigende Stellen hingewiesen (z.B. Regurgitation).
Similar content being viewed by others
Literatur
Müller-Mohnssen, H.: Die Strömungsverhältnisse in den Coronararterien usw. Probleme der Coronardurchblutung; Verlag Springer, Göttingen 1958. S. 179
Rudinger, G.: Shock waves in mathematical models of the aorta. J. appl. Mechan.37, Trans. ASME (1970), S. 34
Lambossy, P.: Aperçu historique et critique sur le problème de la propagation des ondes dans un liquide compressible enfermé dans un tube elastique. Helv. Physiol. Acta8, (1950) 209 und9, 145 (1951)
Rosenberg, H.: Instationäre Strömungsvorgänge in Leitungssystemen mit flexibel-elastischen Rohrwänden. Diss. TH Aachen 1963; Forsch.-Ber. Nordrhein-Westfalen Nr. 1450
Lighthill, M.J.: Vicosity effects in sound waves of finite amplitudes. in G.K. Batchelor, H. Bondi: Survey in Mechanics. Cambridge Univ. Press 250 (1956)
Zeller, H., Talukder, N., Lorenz, J.: Model studies of pulsating flow in arterial branches and wave propagation in blood vessels. Agard Conf. Proc.65, 15 (1970). — Versuche zur pulsierenden Strömung durch arterielle Verzweigungen und zur Wellenausbreitung in großen Blutgefäßen. Abhdlgn. Aerodyn. Inst. Aachen20, 41 (1971)
Ravindran, R.: Wellenausbreitung durch nicht-Newtonsche Flüssigkeiten in elastischen Leitungen. Dissert. TH Aachen 1973
Lorenz, J.: Über die Wellenausbreitung in viscoelastischen Leitungen und ihre gasdynamische Analogie. Diss. TH. Aachen 1973
Teipel, J.: Nichtlineare Wellenausbreitungsvorgänge in elastischen Leitungen. Acta Mech.16, 93 (1973)
Zeller, Lorenz, J.: Analogous treatment of wave propagation in liquid-filled elastic tubes and gas-filled rigid tubes. Euromech. Coll. London 1973
Dropmann, K.: Über die Prädilektionsstellen der Atherosklerose und die hämodyn. Verhältnisse in Krümmerstrecken und an Teilungsstellen. Zs. f. Kreislaufforschg.52, 171 (1963)
Talukder, N., Zeller, H.: Modellversuche zur Strömung in arteriellen Verzweigungen. Dt. Ges. Biomed. Technik Hannover 1974. S. 163
Talukder, N.: Untersuchung zur Strömung in arteriellen Verzweigungen. Diss. TH. Aachen 1974
Fry, D.L.: Acute vascular endothelic changes associated with increased blood velocity gradients. Circulation research22, 165 (1968)
Caro, C.G., Fitzgerald, J.M., Schröter, R.C.: Atheroma and arterial wall shear. Proc. Roy. Soc. London B177, 109 (1971)
Gerlach, R., Lymberopoulos, S.: Modellstudien zur Strömungsführung in verschiedenen Gefäßanastomosen. Dt. Ges. Biomed. Techn. 1974. S. 159. Strömungstechn.-vergleichende Untersuchungen innerer arteriovenöser Anatomosen. Dt. Ges. Biomed. Techn. 1973. S. 101
Reul, H.: Funktions- und Strömungsuntersuchungen an natürlichen und künstlichen Herzklappen usw. Diss. TH. Aachen 1971
Chmiel, H.: Zur Blutrheologie in Medizin und Technik. Habilitationsschrift TH. Aachen 1973
Kalmar, P.: Künstliche Herzklappen. Dt. Ges. Biomed. Techn. Erlangen 1973. S. 207
Kramer, C., Naumann, A., Bleifeld, W.: Effert, S.: Über die Durchströmung künstlicher Herzklappen. Thoraxchirurgie, Vasculäre Chirurgie16, 269 (1968)
Naumann, A.: Strömungsfragen der Medizin. Arb.-Gem. f. Forschg. Nordrh.-Westf. H. 203 (1969), S. 27
Naumann, A.: Über Strömungsvorgänge im menschlichen Körper. Acta medicotechnica18, 46 u. 59 (1970)
Naumann, A., Kramer, C.: Flow investigations on artificial heart valves. AGARD Conf. Proc.65, 4 (1971)
Belhouse, B.J., Talbot, L.: The fluid mechanics of the aortic valve. J. Fluid Mech.35, 721 (1969)
Kramer, C.: Künstliche Herzklappen und mechanische Hämolyse. Habilitationsschr. TH. Aachen 1973
Köhler, J., Kramer, C.: Zum Schließvorgang künstlicher Herzklappen. Acta Medicotechn.20, 36 (1972)
Störmer, B., Mendling, W., Kremer, K., Köhler, J.: Vergleichende Modelluntersuchungen zu künstlichen Herzklappen im Kreislaufsimulator. Dt. Ges. f. Thorax-, Herz- u. Gefäßchirurgie Bad Nauheim 1974
Köhler, J., Lambert, J., Heuser, G.: Strömungsuntersuchungen zum Füllvorgang einer pulsierend fördernden Blutpumpe. Dt. Ges. Biomed. Techn. 1973. S. 121
Köhler, J., Blumenfeld, M., Eckstein, H., Heuser, G.: Modelluntersuchungen zur Strömung in Pumpen und Ventilen zur Blutförderung. Dt. Ges. Biomed. Techn. 1974. S. 165
Köhler, J., Heuser, G.: Verbesserung der Kammerfüllung einer pulsierend fördernden Blutpumpe mit künstlichem Vorhof. Abhdlgn. Aerodyn. Institut TH. Aachen, H. 21 (1974), S. 42
Gehrmann, G., Bleifeld, W., Loogen, F.: Mechanische Hämolyse nach Implantation künstlicher Herzklappen. Zs. f. Kreislaufforschg.55, 25 (1966)
Gehrmann, G., Loogen, F.: Mechanische hämolytische Anämie nach Implantation künstlicher Aortenklappen. Dt. Mediz. Wochenschr.89, 625 (1964)
Kramer, C.: Studies on flow induced mechanical hemolysis. AGARD Conf. Proc.65, 5 (1970)
Kramer, C., Sand, P., Bleifeld, W.: Blutströmung und mechanische Hämolyse. Zs. f. Biomed. Techn.16, 164 (1971)
Bleifeld, W., Kramer, C., Pelzer, H.: Strömungsdynamische Einflußgrößen der mechanischen Hämolyse. Thoraxchirurgie. Vaseuläre Chirurgie21, 130 (1973)
Gerhardt, H.J., Kramer, C.: Investigations of damping in turbulent blood flow. Euromech. Coll. London 1973
Holzhüter, H., Angelkort, B., Kenzel, E., Lambert, J., Pelzer, H., Köhler, J.: Neues Meßverfahren zur quantitativen Erfassung der Thrombozyten — Destruktion in extrakorporalen Kreisläufen. Dt. Ges. Biomed. Techn. 1974. S. 177
Nevaril, C.G., Hellums, J.D., Alfrey, C.P., Lynch, E.C.: Physical effects in red blood cell trauma. J. Amer. Inst. Chem. Eng.15, 707 (1969)
Williams, A.R.: Shear-induced fragmentation of human erythrocytes. Biorheology10, 303 (1972)
Taylor, G.J.: The formation of emulsions in definable fields of flow. Proc. Roy. Soc. A146, 501 (1934)
Schmid-Schönbein, H., Wells, R.: Fluid drop-like transition of erythrocytes under shear. Science18, 288 (1969)
Williams, A.R.: Viscoelasticity of the human erythrocyte membrane. Biorheology11, 313 (1973)
Rand, R.: Mechanical properties of the red blood cell membrane. Biophys. Journ.4, 308 (1964)
Blackshear, P.L.: Mechanical hemolysis in the flowing blood; in Y.C. Fung, N. Perrone, M. Anliker: Biomechanics. Prentice Hall 1972. S. 501.
Benner, K.U., Frede, K.E., Lauterjung, K.L.: In vitro investigation with a new method for continuous and quantitative measurements. Pflügers Arch.310, 142 (1970)
Lambert, J., Egberts, M., Holzhüter, H., Wenzel, E., Faude, N.: Ermittelung der zeitlichen Druckänderung und Blutschädigung im Eingriffsbereich einer Rollenpumpe. Dt. Ges. Biomed. Techn. 1974 S. 173
Lambert, J., Naumann, A.: Zur Strömungsmechanik und Blutschädigung in Hochdruckrollenpumpen. Abhdlgn. Aerodyn. Institut Aachen, H.21, 48 (1974)
Heuser, G., Köhler, J., Lambert, J., Bleifeld, W.: Ein Modell für Detailstudien zur Strömungsmechanik und Blutschädigung im Ballonbereich bei Anwendung der intraaortalen Ballonpulsation. Dt. Ges. Biomed. Techn. 1974 S. 173
Dintonfass, L., Yu, J.S., Grace, C.: Dynamic blood coagulation in Theoretical and Clinical Hemorheology. Springer Verlag 1971. S. 303
Schneider, M.D.: Nature of thrombus in a hollow fiber oxygenator. Amer. Journ. Vet. Res.34, 151 (1973)
Turina, M.: Die Rolle des Blutfilters in der Verhütung des Lungenschadens nach Ganzkörperperfusion. Thoraxchirurgie20, 122 (1972)
Author information
Authors and Affiliations
Additional information
Vortrag auf der 108. Versammlung der Gesellschaft Deutscher Naturforscher und Ärzte, Berlin, 16.–19. September 1974.
Ord. em. Professor für Strömungslehre. Herrn Dipl.-Ing. J. Lambert sei für die Unterstützung bei der Abfassung dieses Berichtes gedankt.
Rights and permissions
About this article
Cite this article
Naumann, A. Strömung in natürlichen und künstlichen Organen und Gefäßen. Klin Wochenschr 53, 1007–1019 (1975). https://doi.org/10.1007/BF01614302
Issue Date:
DOI: https://doi.org/10.1007/BF01614302