Abstract
A prototype electronic monitoring stethoscope was constructed from readily available, high-quality components. It consisted of a conventional precordial or esophageal probe connected to a microphone by a rubber adapter. The microphone was connected by lightweight wire to an amplifier and headphones. Twenty-one anesthesia clinicians evaluated the stethoscope and responded to a multiple-choice preference questionnaire. The electronic stethoscope was judged to perform better than the conventional stethoscope in most categories evaluated. The electronic device was perceived to be louder, clearer in sound reproduction, more efficacious for monitoring, and easier to use continuously, and its head-phones were considered more comfortable than the conventional carpiece. Based on our results, we conclude that amplified stethoscopes have the potential to improve monitoring. Further development of electronic stethoscope monitoring seems warranted and is continuing.
Similar content being viewed by others
References
Cushing H. Some principles of cerebral surgery. JAMA 1909;L11:184–192
Anand JS. A simple pediatric constant-monitoring device. Anesth Analg 1973;51:387–388
Patterson JF. Stethoscope monitoring during anesthesia. Anesth Analg 1966;45:572–575
Ploss RE. A simple constant monitor system. Anesthesiology 1955;16:466–467
Ploss WR. An acoustic switch for use in constant monitoring. Anesthesiology 1968;29:1052–1054
Nelson DS. A monaural stethoscope for anesthesiologists. Anesth Analg 1972;51:177–179
Weeks DB. An earphone for use with Doppler blood-pressure monitoring units. Anesth Analg 1974;53:343–344
Newbower RS, Cooper JB, Long CD. Failure analysis—the human element. In: Gravenstein JS, Newbower RS, Ream AK, Smith NT, eds. Essential noninvasive monitoring in anesthesia. New York: Grune and Stratton, 1980:269–281
Cooper JB, Long CD, Newbower RS, Philip JH. Critical incidents associated with intraoperative exchanges of anesthesia personnel. Anesthesiology 1982;56:456–461
Abelson D. A high fidelity electronic stethoscope. JAMA 1971;218:741
deLeon AC, Harvey WP, Canfield D, Johnson E. Group teaching of auscultation: use of a new wireless stethoscope-type headphone. Am J Cardiol 1978;41:333–335
Douglas R, Doddapanei B, Esophageal stethoscope amplifier. Anesth Analg 1984;64:377–378
Colton T. Statistics in medicine. Boston: Little, Brown, 1974:219–221
Ed-Sci Development. Ed-sci statistics. San Francisco: Ed-Sci Development, 1981
Documenta Geigy. Scientific tables, 6th ed. Ardsley, NY: Geigy Pharmaceuticals, 1969:128
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Philip, J.H., Raemer, D.B. An electronic stethoscope is judged better than conventional stethoscopes for anesthesia monitoring. J Clin Monitor Comput 2, 151–154 (1986). https://doi.org/10.1007/BF01620547
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/BF01620547