Abstract
A survey was conducted to determine the accuracy and quality control of automated haematology analysers used in non-clinical toxicity studies. Pooled blood samples from male Sprague-Dawley rats were distributed to 98 laboratory facilities throughout Japan, the samples being delivered under refrigeration to each facility within 18 h of sample preparation. At each facility, the samples were analysed within 4 h of receipt. Commercially available normal human blood samples from a single lot were also analysed at the same time. Most haematological results were within the mean ±3SD (standard deviation), but some facilities gave either high or low values consistently for both human and rat samples. No facility gave high or low values to certain parameters sporadically, which suggests no problem with the accuracy of the equipment. However, it was suspected that there would be some problem in comparing analytical values determined in a unique way by specific equipment design. The use of certain equipment resulted, in rat haematocrit values in particular, being either too high or too low. In these cases, it was deemed necessary to make some adjustments or calibration changes. There were also platelet values with a ‘plus drift’ which was apparently due to contamination with, or failure to identify small red blood cells (RBC). There was no deviation in values which could be attributed solely to the mechanical operation of any of the analytical equipment.
Non-standard, initial setting up of the equipment (originally intended for human use, but now used for a variety of animal species) has been recognised as the main cause for a wider range of the analytical values seen. The results of this survey suggest that it may be necessary to review equipment calibration at each facility, and to re-establish the historical background data.
Similar content being viewed by others
References
Davies DT, Fisher GV (1991) The validation and application of the Technicon H*1 for the complete automated evaluation of laboratory animal hematology. Comp Haematol Intl 1:91–105
Drewinko B, Bollinger P, Rountree M et al. (1982) Eight-parameter automated hematology analyzers: comparison of two flow cytometric systems. Am J Clin Pathol 78:738–747
Foster HS (1990) Governmental quality assurance regulation of haematology laboratories. Clin Lab Haematol 12:Suppl 1, 143–155
Hayashi Y, Matsuzawa T, Unno T et al. (1995) Effects on hematology parameters during cold storage and cold transport of rat and dog blood samples. Comp Haematol Int 5:251–255
Jones GR, Faust MA, Matthews AR (1995) Quality team approach in evaluating three automated hematology analyzers with five-part differential capability. Am J Clin Pathol 103:159:166
Klee GG (1990) Performance goals for internal quality control of multichannel haematology analysers. Clin Lab Haematol 12:Suppl 1,65–74
Leblanc A, Goguel A (1990) The French national system for external quality assessment. Clin Lab Haematol 12: Suppl 1, 129–134
Lippi U, Schinella M, Modena N et al. (1990) Advantages of a new anticoagulant in routine haematology of the Coulter Counter S-Plus STKR analyzer. Am J Clin Pathol 93:760–764
Matsuzawa T, Nomura M, Unno T (1993) Clinical pathology reference ranges of laboratory animals. J Vet Med Sci 55:351–362
Matsuzawa T, Tabata T, Sakazume M et al. (1994) A comparison of the effect of bleeding site on haematological and plasma chemistry values of F344 rats: the inferior vena cava, abdominal aorta and orbital venous plexus. Comp Haematol Int 4:207–211
Matsuzawa T, Nomura M, Yonezawa H et al. (1995) Selection of appropriate parameters, use of a quality control concept and suitable statistical analyses for clinical pathology examination of animals in toxicity studies: results if a current survey by the Japanese Pharmaceutical Manufacturers Association. Comp Haematol Int 5:196–200
Reardon DM, Hutchinson DD, Preston FFE et al. (1985) The routine measurement of platelet volume: a comparison of apertureimpedance and flow cytometric systems. Clin Lab Haematol 7:251–257
Riley JH (1992) Clinical pathology: preanalytical variation in preclinical safety assessment studies — Effect on predictive value of analyte tests. Toxicol Pathol 20:Part 2, 490–500
Robertson EP, Lai HW, Wei, DCC (1992) An evaluation of leucocyte analysis on the Coulter STKS. Clin Lab Haematol 14:53–68
Rustand P, Lund KP (1990) Practical experience with a quality control procedure using retained patient specimens on Technicon Hl and Coulter S 880, Eur J Haematol 45, suppl 53:19–21
Smith CN, Neptun DA, Irons RD (1986) Effect of sampling site and collection method on variations in baseline clinical pathology parameters in Fischer-344 rats. II. Clinical haematology. Fund Appl Toxicol 7:658–663
Tabata H, Nakamura S, Matsuzawa T (1995) Some species differences in the false prolongation of prothrombin times and activated partial thromboplastin times in toxicology. Comp Haematol Int 5:140–144
Tatsumi N, Tsuda I, Kondoh H (1993) Standardization in hematology and the use of the data in clinical laboratories. Jpn J Clin Pathol 41:592–595
Unno T, Morita N, Saito M et al. (1994) Maintenance and accuracy control of various instruments for clinical pathology examination in the toxicity study. J Toxicol Sci (Jpn) 19:338
Watanabe K, Kawai Y, Takeuchi K et al. (1991) Difference of blood cell counts with reference blood cell counters in three different makers. Rinsho-Byori (Jpn J Clin Pathol) 39:1109–1112
Weiser MG (1983) Comparison of two automated multi-channel blood cell counting systems for analysis of blood of common domestic animals. Vet Clin Pathol 12:25–32
Weigand KW, Odioso LW, Dameron GW et al. (1992) Hematology analyzer comparison: orotho ELT-8/ds vs Baker 9000 for healthy dogs, mice and rats. Vet Clin Pathol 21:10–14
Wolford ST, Schroer RA, Gobs FX et al. (1986) Reference range data for serum chemistry and haematology values in laboratory animals. J Toxicol Environ Health 18:161–188
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Matsuzawa, T., Morita, N., Hayashi, Y. et al. A survey of the results of haematological parameters, using a common rat blood sample in japanese laboratories. Comp Haematol Int 6, 125–133 (1996). https://doi.org/10.1007/BF00368455
Issue Date:
DOI: https://doi.org/10.1007/BF00368455