Abstract

We demonstrate that neutron fluences of nearly 1000 cm−2 can be measured when the fonctions of an activation detector and a device for recording induced activity are combined in one object i.e., a NaI(Tl) scintillation crystal. The beta radiation of the product from the127I(n, γ)128 reaction is detected with close to 100% efficiency by the crystal. The detector can be used in work with low-intensity sources of neutron fields such as neutron generators, radiosotope sources, and TOKAMAK devices.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. V. D. Sevast'yanov, Izmer. Tekh., No. 2, 50 (1990).

    Google Scholar 

  2. K. Bekurtz and L. Wirtz, Neutron Physics [Russian translation], Atomizdat, Moscow (1968).

    Google Scholar 

  3. Handbook of Beta Radiation of Fission Products [in Russian], Atomizdat, Moscow (1978).

Download references

Authors

Additional information

Translated from Izmeritel'naya Tekhnika, No. 2, p. 56, February, 1995.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Antropov, S.Y., Ermilov, A.P. High-sensitivity activation neutron detector. Meas Tech 38, 227–228 (1995). https://doi.org/10.1007/BF00979641

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00979641

Keywords

Navigation