Abstract
A simple and portable apparatus was developed for measurements of85Kr and133Xe that would be released into the atmosphere in an emergency situation of nuclear facilities. The method is based on cryogenic adsorption of these gases on charcoal followed by chromatographic separation from other gases. The85Kr and133Xe recovered from atmospheric air are determined separately by liquid scintillation counting. It takes about 1 hour for the stepwise determination of85Kr and133Xe. The atmospheric concentration of 3·10−3 μCi per m3 air (1.1·102 Bq/m3 air) is measurable for both nuclides with 20% counting error.
Similar content being viewed by others
References
P. PLATO, C. HUDSON, Health Phys., 38 (1980) 523.
K. KIGOSHI, Bull. Chem. Soc. Japan, 35 (1962) 1014.
R. SHUPING, C. PHILLIPS, A. MOGHISSI, Anal. Chem., 41 (1969) 2082.
D. STEVENSON, F. JOHNS, IAEA-SM-148, 1971, p. 157.
S. CUMMINGS, R. SHEARIN, C. PORTER, IAEA-SM-148, 1971, p. 163.
F. PATTI, P. BOURGEON, J. Radioanal. Chem. 56 (1980) 221.
T. OKAI, Y. TAKASHIMA, N. SHIRAISHI, N. MATSUOKA, J. Radioanal. Nucl. Chem., 81 (1984) 161.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Matsuoka, N., Momoshima, N., Okai, T. et al. An improved apparatus for the determination of krypton-85 and xenon-133 in an emergency situation. Journal of Radioanalytical and Nuclear Chemistry, Articles 120, 21–27 (1988). https://doi.org/10.1007/BF02037847