Objective
To analyze efficiency and reliability of 4 modern neonatal ventilators under difficult test conditions. The ventilators tested were: Babylog 8000 (Dräger Medical), BP 2001 (Bear Medical Systems), Sechrist IV 100 B (Sechrist Industries), Infant Star (Infrasonics INC).
Measurements and results
Gas flow generation was tested by comparison of preset flow values with no resistance in the circuit to flow values obtained during interposition of a resistance in the inspiratory circuit. A decrease in gas flow was observed when interposition of a resistance in the inspiratory circuit increased peak inspiratory pressure to 60 cmH2O (gas flow decreased by 8% to 24% depending on the ventilator tested). The pressure limiting valve and the positive end-expiratory pressure valve were also evaluated in order to test their behaviour under different flow conditions. Flow-dependence of the pressure was noted for all ventilators except Babylog 8000. Assessment of the reliability of pressure monitoring revealed either ‘under’ or ‘over’ estimation of peak inspiratory pressure and positive end-expiratory pressure depending on the ventilator tested.
Conclusion
For the best clinical use of mechanical ventilators, neonatologists should be aware of these limitations. Therefore a regular assessment of ventilator performance and monitoring reliability is recommended.
Article PDF
Similar content being viewed by others
References
Simbruner G, Gregory G (1981) Performances of neonatal ventilators: the effects of changes in resistance and compliance. Crit Care Med 9:509–514
Nunn JF (1986) Bench testing of the CPU-1 ventilator. Br J Anesth 58: 653–662
Martin LD, Rafferty JF, Wetzel RC, Gioia FR (1989) Inspiratory work and response times of a modified pediatric volume ventilator during synchronized intermittent mandatory ventilation and pressure support. Anesthesiology 71: 977–981
Fredberg JJ, Glass GM, Boyton BR, Frantz ID (1987) Factors influencing mechanical performance of neonatal high-frequency ventilators. J Appl Physiol 62:2485–2490
Peslin R (1969) The physical properties of ventilators in the inspiratory phase. Anesthesiology 30:315–324
Carlo WA, Martin RJ (1983) Principles of neonatal assisted ventilation. Pediatr Clin North Am 10:221–237
Cunningham D, Desai N (1986) Methods of assessment and findings regarding function in infants less than 1000 grams. Clin Perinatol 13:299–313
Kirpalani H, Santos-Lyn R, Roberts R (1988) Some infant ventilators do not limit peak inspiratory pressure reliability during active expiration. Crit Care Med 16:880–883
Marini JJ, Culver B, Kirk W (1985) Flow resistance of exhalation valves and positive end-expiratory pressure devices used in mechanical ventilation. Am Rev Respir Dis 131:850–854
Flynn JT (1987) Retinopathy of prematurity. Pediatr Clin North Am 34: 1487–1516
Beydon L, Liu N, Hassapopoulos J, Rua F, Bonnet F, Harf A, Duvaldestin P (1992) Test of similar intensive care ventilators in daily use conditions. Evaluation of accuracy and performances. Intensive Care Med 18:32–37
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Jouvet, P., Hubert, P., Jarreau, P.H. et al. Assessment of neonatal ventilator performances. Intensive Care Med 21, 753–758 (1995). https://doi.org/10.1007/BF01704743
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01704743