Skip to main content
Log in

Interference by cephalosporins with creatinine measurement by desk-top analyzers

  • Short Communications
  • Published:
European Journal of Clinical Pharmacology Aims and scope Submit manuscript

Summary

We have carried out a study to evaluate the interference by cephalosporins with the measurement of creatinine by desk-top analyzers. The cephalosporins evaluated at concentrations of 0–250 mg/l were cefazolin sodium, cefoxitin sodium, cefotaxime sodium, and ceftazidime pentahydrate. The instruments evaluated were DT60 (Kodak, Rochester, USA), Seralyzer (Ames Division, Miles Laboratories, IN, USA), and Vision (Abbott Labs, Chicago, USA). All studies were done in plasma.

None of the cephalosporins showed any interference with the DT60 analyzer. With the Vision and Seralyzer no interference was seen with cefotaxime or cefazolin. With cefazolin an increase of 10–20 µmol/l creatinine was seen for every 20 mg/l of drug; with cefoxitin there was an increase of 50–80 µmol/l of creatinine for every 100 mg/l of drug.

Erroneous creatinine values may be found in patients taking cefazolin and cefoxitin and may lead to inappropriate clinical management.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

References

  1. Swain RR, Briggs SL (1977) Positive interference with the Jaffé reaction by cephalosporin antibiotics. Clin Chem 23: 1340–1342

    Google Scholar 

  2. Saah AJ, Kock TR, Drusano GL (1982) Cefoxitin falsely elevates creatinine levels. J Am Med Assoc 247: 205–206

    Google Scholar 

  3. Letellier G, Desjarlais F (1985 Analytical interference of drugs in clinical chemistry. Clin Biochem 18: 352–356

    Google Scholar 

  4. Hyneck ML, Berardi RR, Johnson RM (1981) Interference of cephalosporins and cefotoxin with serum creatinine determination. Am J Hosp Pharm 38: 1348–1352

    Google Scholar 

  5. Belsey R, Baer D, Sewell D (1986) Laboratory test analysis near the patient. J Am Med Assoc 255: 775–786

    Google Scholar 

  6. Marks V (1983) Clinical biochemistry nearer the patient. Br Med J 286: 1166–1167

    Google Scholar 

  7. Kroll MH, Elin RJ (1983) Mechanism of cefotoxitin and cephalothin interference with the Jaffé method for creatinine. Clin Chem 29: 2044–2048

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Nanji, A.A., Poon, R. & Hinberg, I. Interference by cephalosporins with creatinine measurement by desk-top analyzers. Eur J Clin Pharmacol 33, 427–429 (1987). https://doi.org/10.1007/BF00637643

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00637643

Key words

Navigation