Summary
We have developed a frameless stereotactic multi-articulated arm suitable for image-directed microsurgery and surgical navigation. The system comprises a frameless isocentric mechanism and a three-dimensional (3 D) digitizer for intra-operative monitoring. Five joints of our multi-articulated arm are equipped with electromagnetic clutches, and the arms are provided with counterbalancing weights to ensure safe, smooth movement. Stereotactic microsurgical procedures were carried out with a laser beam guide mounted at the tip of the isocentric arm.
We have used this system in 22 cases to resect tumours or epileptic foci. There were 12 cases of glioma, 5 cases of deep seated benign tumours, 3 cases of multiple metastatic tumours, and 2 cases of temporal lobe epilepsy. Postoperative enhanced CT images indicated total or subtotal resection in all cases. We also have used this system to determine the approach for clipping of a posterior cerebral artery aneurysm in 2 cases. The error of mechanical accuracy of the system was less than 1 mm. We believe that this system will prove useful in image-directed microsurgical procedures.
Similar content being viewed by others
References
Guthrie BL, Kaplan R, Laws ER Jr (1990) Three-dimensional cranial image reconstruction for interactive freehand stereotaxy. In: Proceedings of the American Association of Neurological Surgeons, pp 357 (Abstract)
Kato A, Yoshimine T, Hayakawa T, Tomita Y, Ikeda T, Mitomo M, Harada K, Mogami H (1991) A frameless armless navigation system for computer-assisted neurosurgery. J Neurosurgery 74: 845–849
Kelly PJ, Kall BA, Goerss SJ, Earnest F (1986) Computer-assisted stereotactic resection of intra-axial brain neoplasm. J Neurosurg 64: 427–439
Laitinen LV, Liliequist B, Fagerlund M, Eriksson AT (1985) An adapter for computed tomography-guided stereotaxis. Surg Neurol 23: 559–566
Leksell L, Linquist C, Alder JR, Leksell D, Jernberg B, Steiner L (1987) A new fixation device for the Leksell stereotactic system. Technical note. J Neurosurg 66: 626–628
Roberts DW, Strohbehn JW, Hatch JF, Murray W, Kettenberger H (1986) A frameless stereotactic integration of computerized tomographic imaging and the operating microscope. J Neurosurg 65: 545–549
Takizawa T (1993) Iso-centric stereotactic three dimensional digitizer for neurosurgery. Stereotact Funct Neurosurg 60: 175–193
Watanabe E, Watanabe T, Manaka S, Mayanagi Y, Takakura K (1987) Three dimensional digitizer (neuro-navigator). A new equipment for CT guided stereotaxic surgery. Surg Neurol 27: 543–547
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Takizawa, T., Soto, S., Sanou, A. et al. Frameless isocentric stereotactic laser beam guide for image-directed microsurgery. Acta neurochir 125, 177–180 (1993). https://doi.org/10.1007/BF01401848