Addition of modifier in supercritical fluid chromatography using a microbore reciprocating pump

  • Originals
  • Published:
Chromatographia Aims and scope Submit manuscript

Summary

Solubilities of polar analytes in supercritical CO2 can be enhanced by the addition of polar modifiers. The addition of modifier into an SFC system using a low cost reciprocating pump has been studied. Two different mixing chambers were evaluated for mixing the supercritical CO2 with modifier. It appeared that a mixing chamber with a packed bed was enough to reduce baseline noise from the modifier pump. Results from the effect of pressure and temperature with various modifier flow rates were obtained. High percentages of modifier (>15%) at a low CO2 pressure (2000 psi) caused baseline instability. In addition, different I.D. columns were tested with the system and the effect of modifier compressiblity on detector noise was also studied. Several pharmaceutical compounds were separated to demonstrate system performance.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. M. Ashraf-Khorassani, S. Shah, L. T. Taylor, Anal. Chem.62, 1173 (1990).

    Google Scholar 

  2. D. R. Gere, R. Board, D. McManigill, Anal. Chem.54, 736 (1984).

    Google Scholar 

  3. M. Ashraf-Khorassani, L. T. Taylor, Elution behavior of nitrogen-containing compounds of different basicity from various stationary phases employing supercritical mobile phases, in “modern supercritical fluid chromatography”, C. M. White, Ed., Hüthig Verlag, Heidelberg, Germany, 1988, Ch. 7.

    Google Scholar 

  4. T. A. Berger, J. F. Dye, Unpublished results.

  5. J. M. Levy, W. M. Ritchey, J. Chromatogr. Sci.24, 242 (1986).

    Google Scholar 

  6. T. A. Berger, J. F. Dye, J. Chromatogr. Sci.29, 390 (1991).

    Google Scholar 

  7. T. A. Berger, J. F. Dye, M. Ashraf-Khorassani, L. T. Taylor, J. Chromatogr. Sci.27, 105 (1989).

    Google Scholar 

  8. M. Ashraf-Khorassani, M. Fessehaie, L. T. Taylor, T. A. Berger, J. F. Dye, J. HRC11, 352 (1988).

    Google Scholar 

  9. T. A. Berger, J. F. Dye, J. Chromatogr. Sci.29, 310 (1991).

    Google Scholar 

  10. T. A. Berger, J. F. Dye, J. Chromatogr. Sci.29, 26 (1991).

    Google Scholar 

  11. T. A. Berger, J. F. Dye., J. Chromatogr. Sci.29, 54 (1991).

    Google Scholar 

  12. T. A. Berger, J. F. Dye, J. Chromatogr. Sci.29, 141 (1991).

    Google Scholar 

  13. T. A. Berger, J. F. Dye, J. Chromatogr. Sci.547, 377 (1991).

    Google Scholar 

  14. M. Lee, K. E. Markides, “Analytical supercritical fluid chromatography and extraction. Chromatography Conference Inc. 1991.

  15. J. Via, L. T. Taylor, F. K. Schweighardt, Anal. Chem.66, 1459 (1994).

    Google Scholar 

  16. Lee Company catalogue.

  17. “CRC Handbook of Chemistry and Physics”, CRC prss, 54th edition, F-12.

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ashraf-Khorassani, M., Levy, J.M. Addition of modifier in supercritical fluid chromatography using a microbore reciprocating pump. Chromatographia 40, 78–84 (1995). https://doi.org/10.1007/BF02274611

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02274611

Key Words

Navigation