Response of BaF2 detectors to photons of 3–50 MeV energy

https://doi.org/10.1016/0168-9002(90)90259-9Get rights and content

Abstract

BaF2 detectors of 20 cm length (10 radiation lengths) and hexagonal cross section (diameter 5.2 cm) were tested using monochromatic photons from the tagged-photon facility at the electron accelerator MAMIA at Mainz. The experimental spectra the deposited energy for a single detector and for an array of seven modules compare very well with the predictions of Monte Carlo shower simulations using the code GEANT3. At high photon energies a significant improvement (more than a factor 2) of the energy resolution is observed for the summed energy spectra as compared to the resolution of one single module.