Summary
Quantitative transmission profile scanning of the middle phalanx of the left forefinger was performed using the method of Börneret al. Normal ranges, different for sex and age, were established by the investigation of 86 patients showing no evidence of bone disease.
The evaluation of this method's reproducibility in 5 normal subjects gave the following results:
-
S. D. at five different days 7–35%
-
S. D. at five continuous determinations 3.5–9.0%.
Of 47 patients with severe chronic renal insufficiency, 33 showed signs of osteoporosis evidenced by radiological examination. Only in 13 of them were the absorption coefficients significantly decreased. Of the 14 patients without radiological evidence of osteoporosis 7 had normal and 7 other decreased values.
It is concluded that the reproducibility of the method is not sufficient for follow-up studies and that the forefinger might not be representative for bone decalcification caused by chronic renal insufficiency.
Zusammenfassung
Die quantitative Bestimmung der Absorption einer125J-Strahlenquelle durch die Mittelphalanx eines Fingers soll die frühzeitige Erkennung und Verlaufskontrolle von Skeletkalksalzminderungen gestatten.
Der vorläufige Normalwertbereich von 86 Probanden deckt sich mit den Angaben in der Literatur, während die Reproduzierbarkeit der Einzel- als auch der wiederholten Messungen deutlich abweicht. Die radiologische Osteoporose bei chronisch niereninsuffizienten Patienten konnte nur in 33% durch die Messung der Absorption am Finger bestätigt werden.
Literatur
Börner, W., Grehn, S., Moll, E., Rauh, E.: Messung der Absorption des Fingerknochens mit einem125J-Profilscanner. Quantitative Methode zur Erkennung der Osteoporose. Fortschr. Röntgenstr.110, 378 (1969).
Börner, W., Heieis, G., Moll, E., Rauh, E., Bracharz, H., Longin, F.: Therapie-Kontrolle bei Osteoporose-Kranken mit einem125J-Profilscanner. Klin. Wschr.47, 1115–1116 (1969).
Grehn, S.: Osteoporosediagnostik mit125J, verbesserte Methode und ein neues Auswerteverfahren für Absorptionsmessungen am Fingerknochen. Inaugural-Dissertation, Würzburg 1968.
Börner, W., Grehn, S., Moll, E., Rauh, E.: A125J- profile scanner for measuring bone density in the fingers. Germ. med. Mth.14, 301–305 (1969).
Börner, W., Grehn, S., Moll, E., Rauh, E., Seybold, K.: Altersphysiologische und pathologische Veränderungen der Dichte und Dicke des Fingerknochens. Radiologishce Messungen mit einem125J-Profilscanner an 223 Frauen. Fortschr. Röntgenstr.116, 552–558 (1972).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Breuel, H.P., Hesch, R.D., Henning, H.V. et al. Osteoporose-Diagnostik mit einem125J-Profilscanner bei renaler Osteopathie. Klin Wochenschr 51, 767–768 (1973). https://doi.org/10.1007/BF01468283
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF01468283