A comparative study of constant-voltage and constant-temperature hot-wire anemometers: Part II – The dynamic response

Part II – The dynamic response

  • Published:
Experiments in Fluids Aims and scope Submit manuscript

Abstract

The dynamic response of the constant-voltage anemometer (CVA) system was investigated both analytically and experimentally and compared to that of the CTA. The frequency response functions of the CVA system for a number of different circuit parameters and flow conditions were determined via laser-based radiative heating of the hot-wire sensor. A 2nd-order linear systems model of the CVA was developed to provide insight to the dynamic response and to interpret the experimental results. The qualitative variations in the frequency response function with changes in circuit parameters are in agreement with the model. The experimentally determined frequency-response functions of the CVA systems used in this study were found to have little dependence on the wire overheat ratio and Reynolds number.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: 19 July 1998/Accepted: 26 April 1999

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kegerise, M., Spina, E. A comparative study of constant-voltage and constant-temperature hot-wire anemometers: Part II – The dynamic response. Experiments in Fluids 29, 165–177 (2000). https://doi.org/10.1007/s003489900074

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s003489900074

Keywords

Navigation