Causes of severe decentration and subluxation of intraocular lenses

  • Clinical Investigation
  • Published:
Graefe's Archive for Clinical and Experimental Ophthalmology Aims and scope Submit manuscript

Abstract

• Background: Severe decentration and subluxation of intraocular lenses (IOLs) may lead to double vision, glare and deterioration of vision to the point of functional aphakia. The purpose of the present study was to analyse causes for severe IOL dislocation. • Material and methods: Between January 1989 and January 1996, 37 patients required IOL exchange because of decentrated or subluxated posterior chamber lenses. Twenty-five of the exchanged lenses were implanted in our hospital, 12 lenses elsewhere. After explantation the lenses were examined by light and electron microscopy. • Results: In 10 eyes, asymmetric implantation of the posterior chamber lens was responsible for decentration. Three of the lenser concerned were multifocal IOLs. Asymmetric implantation led to a significantly higher rate of explantations in eyes with multifocal lenses (P<0.005). In five eyes decentration developed due to asymmetric capsular shrinkage, in four eyes due to posterior synechiae. A lens subluxation developed in three eyes as a result of rupture of the posterior capsule and in nine eyes because of zonular defects. In three cases decentrations were induced by an extensive secondary cataract. Macroscopically visible changed geometry of the haptics was found in nine lenses; eight of these had polypropylene haptics. Seven lenses showed severely altered haptics on electron-microscopic examination. In four eyes subluxated lenses had to be explanted together with the capsular bag because of severe defects of the zonula, which caused decentration in nine eyes. • Conclusions: Asymmetric implantation of posterior chamber lenses should be strictly avoided. Multifocal lenses require special attention concerning symmetric capsulorhexis and positioning of their haptics.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Anders N, Pham DT, Walkow T, et al. (1996) IOL-Austausch und Explantation — eine retrospektive Analyse. Vorösmarthy D (ed): 10. Kongreß der deutschsprachigen Gesellschaft für Intraokularlinsenimplantation und refraktive Chirurgie, Springer, Heidelberg, pp 238–243

    Google Scholar 

  2. Assia RI, Legier UFC, Merril C, et al. (1993) Clinicopathologic study of the effect of radial tears and loop fixation on intraocular lens decentration. Ophthalmology 100: 153–158

    PubMed  CAS  Google Scholar 

  3. Auffahrt GU, Wesendahl ThA, Brown SJ, Apple DJ (1994) Gründe für die Explantation von Hinterkammerlinsen. Ophthalmologe 91: 507–511

    Google Scholar 

  4. Caballero A, Lopez MC, Losada M, et al. (1995) Long-term decentration of intraocular lenses implanted with envelope capsulotomy and continuous curvilinear capsulotomy: a comparative study. J Cataract Refract Surg 21: 287–292

    PubMed  CAS  Google Scholar 

  5. Doren GS, Stern GA, Driebe WT (1992) Indications for and results of intraocular lens explantation. J Cataract Refract Surg 18: 79–85

    PubMed  CAS  Google Scholar 

  6. Frezzotti R, Caporossi A, Mastrangelo D, et al. (1990) Pathogenesis of posterior capsular opacification. II. Histopathological and in vitro culture findings. J Cataract Refract Surg 16: 353–360

    PubMed  CAS  Google Scholar 

  7. Hansen SO, Tetz MR, Solomon KD, et al. (1988) Decentration of flexible loop posterior chamber intraocular lenses in a series of 222 postmortem eyes. Ophthalmology 95: 344–349

    PubMed  CAS  Google Scholar 

  8. Kimura W, Kimura T, Sawada T, et al. (1993) Comparison of shape recovery ratios of single-piece poly(methylmethacrylate) intraocular lens haptics. J Cataract Refract Surg 19: 635–639

    PubMed  CAS  Google Scholar 

  9. Mamalis N, Crandall AS, Pulsifer MW, et al. (1991) Intraocular lens explantation and exchange. J Cataract Refract Surg 17: 811–818

    PubMed  CAS  Google Scholar 

  10. Masket S(1993) Postoperative complications of capsulorhexis. J Cataract Refract Surg 19: 721–724

    PubMed  CAS  Google Scholar 

  11. Ohmi S, Uenoyama K (1995) Decentration associated with asymmetric capsular shrinkage and intraocular lens design in a rabbit model. J Cataract Refract Surg 21: 293–296

    PubMed  CAS  Google Scholar 

  12. Price FW, William WE, Collins K, Johns S (1992) Changing trends in explanted intraocular lenses: a single center study. J Cataract Refract Surg 18: 470–479

    PubMed  Google Scholar 

  13. Sinskey RM, Amin P, Stoppel JO (1993) Indications for and results of a large series of intraocular lens exchanges. J Cataract Refract Surg 19: 68–71

    PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Walkow, T., Anders, N., Pham, D.T. et al. Causes of severe decentration and subluxation of intraocular lenses. Graefe’s Arch Clin Exp Ophthalmol 236, 9–12 (1998). https://doi.org/10.1007/s004170050035

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s004170050035

Keywords

Navigation