Abstract
Background: Skin burns and ignition of drapes have been reported with the use of cold light sources. The aim of the study was to document the temperature generated by cold light sources and to correlate this with the total radiated power and infrared output.
Methods: The temperature, total radiated power, and infrared output were measured as a function of time at the end of the endoscope (which is inserted into the operative field) and the end of the fiber optic bundle of the light cable (which connects the cable to the light port of the endoscope) using halogen and xenon light sources.
Results: The highest temperature recorded at the end of the endoscope was 95°C. The temperature measured at the optical fiber location of the endoscope was higher than at its lens surface (p < 0.0001). At the end of the fiber optic bundle of light cables, the temperature reached 225°C within 15 s. The temperature recorded at the optical fiber location of all endoscopes and light cables studied rose significantly over a period of 10 min to reach its maximum (p < 0.0001) and then leveled off for the duration of the study (30 min). The infrared output accounted only for 10% of the total radiated power.
Conclusions: High temperatures are reached by 10 min at the end of fiber optic bundle of light cables and endoscopes with both halogen and xenon light sources. This heat generation is largely due to the radiated power in the visible light spectrum.
Similar content being viewed by others
Author information
Authors and Affiliations
Additional information
Received: 10 March 1997/Accepted: 1 August 1997
Rights and permissions
About this article
Cite this article
Hensman, C., Hanna, G.B., Drew, T. et al. Total radiated power, infrared output, and heat generation by cold light sources at the distal end of endoscopes and fiber optic bundle of light cables. Surg Endosc 12, 335–337 (1998). https://doi.org/10.1007/s004649900665
Published:
Issue Date:
DOI: https://doi.org/10.1007/s004649900665