Zusammenfassung
Bericht über einen Fall mit rezidivierendem Aneurysma der A. carotis interna — communicans posterior nach erfolgreicher Verclippung mit einem Heifetz-Clip. Ursache des Rezidivs war das Abbrechen eines Clipblattes an seinem proximalen Ende längere Zeit nach der Operation. Ursache scheinen strukturelle Fehler des Aneurysmaclipmetalles gewesen zu sein.
Summary
Recurrence of an IC—PC aneurysm after a successful neck clipping (Heifetz' clip) is reported. This was caused by breaking of a clip blade at its proximal end. It is suggested that some structural weakness was present in the clip metal.
References
Drake, C. G., Venderlinden, R. G., The late consequences of incomplete surgical treatment of cerebral aneurysms. J. Neurosurg.27 (1967), 226–238.
—, Allcock, J. M., Postoperative angiography and the “slipped” clip. J. Neurosurg.39 (1973), 83–689.
Heifetz, M. D., Technical suggestion. A new intracranial aneurysm clip. J. Neurosurg.30 (1969), 753.
Mayfield, F. H., Kees, Jr., G., A brief history of the development of the Mayfield clip. J. Neurosurg.35 (1971), 97–100.
Scoville, W. B., Technical suggestion. Miniature torsion bar spring aneurysm clip. J. Neurosurg.25 (1966), 97.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Hayakawa, I., Sasaki, A., Fujiwara, K. et al. Ein ungewöhnliches Aneurysmarezidiv nach Verclippung — ein struktureller Fehler des Clips?. Acta neurochir 35, 271–277 (1976). https://doi.org/10.1007/BF01406123