A number of capillary columns were prepared by filling them with supports of different mesh size and coating them with different amounts of stationary phase.
The permeability of these columns is ten times higher than that of the conventional packed capillary columns prepared with the same support particles. This explains partly their good performances.
The packed capillary columns thus prepared had a very high efficiency, providing up to 2500 plates per meter for n-heptane at 50°C (k=2.2) and a high optimum speed.
By using serveral different carrier gases, the variation of the height of a theoretical plate with the carrier gas flow rate was determined. The experimental results cannot be interpreted by means of the classical Van Deemter equation; this is a systematic trend which might be the result of a coupling effect (17, 18).
Zusammenfassung
Es wurde eine Anzahl gepackter Kapillarsäulen hergestellt, die mit Trägermaterial verschiedener Korngrösse gefüllt und mit verschiedenen Phasenverhältnissen versehen wurden.
Die Permeabilität dieser Säulen ist zehnmal höher als diejenige von üblichen, mit Trägermaterial gleicher Korngrösse gefüllten Kapillaren; dies ist einer der Gründe für ihre gute Leistung.
Es wurden gepackte Trennkapillaren mit sehr hoher Leistung hergestellt, die bis zu 2500 Böden je Meter für n-Heptan bei 50°C aufwiesen (k=2.2) und die einen hohen optimalen Gasvolumenfluß hatten.
Mit Hilfe verschiedener Trägergase wurde die Änderung der Höhe eines theoretischen Bodens mit dem Trägergasvolumenfluß bestimmt. Die experimentellen Ergebnisse können nicht mit Hilfe der klassischen Van-Deemter-Gleichung ausgewertet werden (17, 18).
Sommaire
On a préparé une série de colonnes capillaires remplies de particules de diamètres variés et imprégnées de différents taux de phase stationnaire.
La perméabilité de ces colonnes est dix fois plus grande que celle des colonnes classiques remplies de particules de même diamètre, ce qui est une des raisons de leurs bonnes performances.
On a pu préparer des colonnes capillaires remplies ayant une efficacité très élevée, jusqu'à 2500 plateaux par mètre pour le n-hep-tane à 50°C (k′=2,2) et une vitesse optimale élevée.
On a déterminé la variation de la hauteur équivalente à un plateau théorique avec la vitesse du gaz vecteur, en utilisant plusieurs gaz vecteurs différents. Les résultats expérimentaux ne peuvent pas être interprétés en utilisant l'équation devan Deemter classique, mais semblent montrer l'existence d'un effet couplé [17, 18].
Similar content being viewed by others
Bibliographie
Halasz, I., Heine, E., Nature194, 971 (1962).
Landault, C., Thèse, Paris, 1967 CNRS no A-0-1321.
Desty, D.M., Geach, C.J., Goldup, A., Gas Chromatography 1960,Scott R. P. W. Butterworths, Londres 1960, p. 46.
Halasz, I., Schreyer, G., Z. Anal. Chem.181, 384 (1961).
Desty, D.M., haresnape, J.H., Whyman, B.H.F. Anal. Chem.32, 302 (1959).
Kreyenbuhl, A., Bull. Soc. Chim. France, (1960) 2125.
McAfee, K.B., (Jr.),in Encyclopiedia of Chemical Technology Ed.Kirk, K.E., Othmer, D.F., The Interscience Encyclopedia. 2ème volume supplémentaire, 1960, p. 308.
Landault, C., Guiochon, G., Bul. Soc. Chim. France., (1963) 2433.
Landault, C., Guiochon, G., Gas Chromatography 1964,Goldup A. Ed. The Institute of Petroleum, London, 1965, p. 121.
Landault, C., Guiochon, G., Anal. Chem.39, 713 (1967).
Landault, C., Guiochon, G., Bull. Soc. Chim. France, 1967
Baker, W.J., Lee, E.H., Wall, R.F. Gas Chromatography,Noebels, H.J., Wall, R.F., Brenner, N., ed. Academic Press, New York 1961, p. 21.
Guiochon, G., Chromatographic Reviews. Lederer, E. ed. Elsevier, Amsterdam 1967, Vol. 8, p. 1.
Hirschfelder, J.O., Curtiss, C.F., Bird, R.B., Molecular Theory of Gases and Liquids. Wiley, New York 1954.
Fuller, E.N., Giddings, J.C., J. Gas Chromatography3, 223 (1965).
Van Deemter, J.J., Zuiderweg, F.J., Klinkenberg, A., Chem. Eng. Sci.5, 271 (1956).
Giddings, J.C., Anal. Chem.35, 439 (1963).
Littlewood, A.B., Anal. Chem.38, 2 (1966).
Halasz, I., J. Gas Chromatography4, 8 (1966).
Knox, J.H., McLaren, L., Anal. Chem.36, 1477 (1964).
Scott, R.P.W., Hazeldean, G.S.F., Gas Chromatography 1960,Scott, R.P.W., ed. Butterworths, Londres 1960 p. 144.
Golay, M.J.E. Gas Chromatography 1958,Desty, D.H., ed. Butterworths, Londres 1958, p. 36.
Chovin, P., Guiochon, G., Bull. Soc. Chim. France, 1965, 3396.
Desty, D.H., Goldup, A., Gas Chromatography 1960,Scott, R.P.W. ed., Butterworths, Londres 1960 p. 162.
Halasz, I., Hartmann, K., Heine, E., Gas Chromatography 1964, A. Goldup ed. The Institute of Petroleum, Londres, 1965, p. 38.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Landault, C., Guiochon, G. Etude théorique et expérimentale des colonnes capillaires remplies utilisées en chromatographie en phase gazeuse I. Etude expérimentale des conditions de fonctionnement. Chromatographia 1, 119–132 (1968). https://doi.org/10.1007/BF02258946
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02258946